Energizer PORTABLE GAS DETECTION INSTRUMENT Instrucciones de operaciones Pagina 41

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 50
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 40
41
SULPHUR DIOXIDE SPECIFICATION
Range:
0.0 to 99.9 ppm
Response Time:
T90 < 35 seconds
Operating Temperature Range:
-20 to +50
o
C
(-4 to +122
o
F)
Graphic Display Temperature Range:
-10 to +60
o
C
(+14 to +140
o
F)
Operating Humidity:
15 to 90% RH, Non-Condensing
Recommended Storage Temperature:
0 to 20
o
C
(32 to 68
o
F)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours - Non-Alarming
About 800 Hours in Battery Save Mode
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: < 3 ppm
Hydrogen Sulfide, 15 ppm: 0 ppm
Nitric Oxide, 35 ppm: 0 ppm
Nitrogen Dioxide, 5 ppm: - 5 ppm
Consult Factory for additional information.