
Alarm
The TWA alarm for toxic gases will operate when the
short-term (15 minute) or long-term (8 hour) time
weighted average alarm levels are exceeded.
TWA alarms cannot be cleared. TWA
alarms are functional for toxic gas instruments only.
Display Menu Options
The G-TECTA™ 4GP menu is displayed
when you press the mode button twice.
There are up to 4 menu options including
the instrument home screen. To scroll
between the different options, press the mode
button once. To select the menu option you require,
double click on the selected icon. When the Peak
and TWA menu options are selected the relevant icon
will be displayed on the screen.
Peak
The Peak menu option displays the highest gas level
for flammable and toxic gases, or the lowest level
of oxygen, since the instrument was switched on or
was last reset.
After 5 seconds the instrument will display Clear
Peak and “????”. Press the mode button to confirm
the Clear Peak function. The instrument will count
down and time out in 10 seconds and change back to
Peak mode if the mode button is not pressed.
The Peak feature is mainly used for confined space
entry checks, where the instrument is lowered into a
potentially hazardous environment (instead of using
a sampling hose), and also for viewing peak gas
exposure at the end of a shift.
The TWA menu option displays the 8 hour time
weighted average reading for toxic gases since the
unit was last switched on.
Zero
The Zero menu option will reset the sensor in the
instrument. When the zero function has been
completed, the instrument will return to normal
operating mode.
Always ensure you are in fresh air
when using this feature. Zeroing an instrument in a
contaminated atmosphere could result in incorrect
gas readings.
Alarms
G-TECTA™ 4GP has two instantaneous alarm level
settings for each sensor – level 1 and level 2 –
and two additional Time Weighted Average (TWA)
alarms for toxic sensors.
Alarm Signals
The red and blue alarm LEDs will flash, the sounder
will emit a loud, fast series of beeps and the internal
vibrator alarm will activate. The display will show the
gas in alarm and the alarm level.
For any gas, there are normally 2 alarm levels*
– these are indicated by the alarm icons:
Note: The G-TECTA 4G is set to “latch” the alarm by
default. This means that the instrument will continue
to alarm until the operator presses the mode button.
Once you have evacuated the dangerous area,
press the mode button on the G-TECTA™ 4GP
instrument; this will reset your G-TECTA™ 4GP
to normal operating mode. If the G-TECTA™ 4GP
continues to alarm, i.e. you are still in a dangerous
area, pressing the button will have no effect.
Alarms
The TWA alarm for toxic gases will operate when
the short-term (15 minute) or long-term (8 hour)
time weighted average alarm levels are exceeded.
The G-TECTA™ 4G will display
and display
toxic gas levels.
TWA alarms are functional for toxic
gas instruments only.
The TWA alarms cannot be cleared.
Flammable over-range alarm
If flammable gas levels exceed 100% LEL,
G-TECTA™
4GP locks into alarm and displays “ ” showing an
over-range condition. G-TECTA™ 4GP will temporarily
cut off power to the sensor to prevent burn out, and
display a progress bar for 200 seconds.
Note: The alarm remains active until the instrument
sensor is in normal operating mode.
When the power timeout is complete, the symbol
changes to “
” and the mode button should be
pressed to put the instrument back into normal
operating mode. The symbol will change to an egg
timer to indicate that the sensor is warming up.
At the end of the warm up period and assuming the
instrument is in fresh air, the sensor will revert to a
numeric display.
Note: If the instrument is still in the hazardous area
when the sensor is re-set, the sensor will revert to
an over-range alarm.
The G-TECTA™ 4GP can optionally be configured
to force the operator to switch the instrument off
and on to restart. This option is programmable with
G-TECTA™ PC Software.
Note: All alarm configurations can be set with
G-TECTA™ PC Software.
If the sensor power is reconnected
when the unit is exposed to an over-range gas
concentration there is a risk of damage to the
pellistor sensor.
Depleted oxygen levels can reduce
the flammable gas reading. If oxygen levels are
below safe breathing limits it should be assumed that
the flammable reading is low.
Instrument Settings
The following instrument settings can be altered
using G-TECTA™ PC Software:
Alarm levels for the sensor:
Level 1 and level 2 alarms can be set for each gas sensor.
Alarm trigger:
The alarm trigger can be set for rising levels of gas
(toxic and flammable), or for falling levels of gas
(oxygen deficiency monitoring).
Alarm latching:
Latched alarms require the operator to press the mode
button to clear the alarm (default setting). The alarm can
be set to be latched or unlatched. Unlatched alarms will
clear automatically when the gas hazard has passed.
*
The instrument has 2 alarm levels to give the user an indication that the gas level may be rising. If any alarm warning is indicated by the
instrument the work area should be evacuated.
CH4 %LELCO ppm
H2S ppm
%
0.0
0.0
0
16.7
O2
Alarm icon
Gas example
under alarm
Comentarios a estos manuales