Gas Test/Calibration Troubleshooting
Problem Possible Cause Solution
No response to gas Gas Cylinder Empty Check cylinder gauge. Replace cylinder
Hose blocked or kinked Check hose and ensure no gas flow restriction
Instrument fails gas test Gas cylinder empty Check cylinder gauge. Replace cylinder
Gas cylinder out of date Check cylinder expiry date. Replace cylinder
Hose blocked or kinked Check hose and ensure no gas flow restriction
Calibration drifted Calibrate instrument
Gas flow not started in time Repeat gas test. Start gas immediately
Instrument fails calibration Gas cylinder empty Check cylinder gauge. Replace cylinder
Gas cylinder out of date Check cylinder expiry date. Replace cylinder
Hose blocked or kinked Check hose and ensure no gas flow restriction
Calibration drifted Calibrate instrument
Instrument stabilisation time
too short
Reset instrument stabilisation time using
G-TECTA™ PC Software
Instrument passes gas test but
will not enter calibration mode
Zero not performed from options menu Initiate zero on instrument menu
Instrument incorrectly configured Re-configure instrument.
Send to Linde Group service centre
G-TECTA™ 4GP Product Specification
Dimensions
122 x 128 x 57 mm
Weight
498g (with attachment clip & 4 sensors)
IP Rating (Dusts & Liquids)
IP65 NEMA 4
Operating Temperature Rating
-20°C Ta +55°C (-4°F Ta +131°F)
Humidity Rating
20-90% RH – non-condensing, continuous operation
Display
128 x 64 Pixel with backlight
Warm-up time period
Approx 45 seconds
Typical response time
T90: 20 seconds (toxic & flammable)
10 seconds (oxygen)
Alarm sounders*
94dB(A)
Optional multiple alarm sounds
Visible alarms
Dual colour Blue/Red flashing LEDs
Vibrating alarm
Internal vibrating
Repeatability
±2% FSD, 6 months
Electromagnetic Compatibility
2004/108/EC
Explosion Protection Certification
IECEx/ATEX/UL/CSA
IECEx
IECEx BAS 09.0118
ATEX
BASEEFA09ATEX0246
UL
UL913
CSA
CSA22.2, 152
Approval Codes Global
IECEx II 2G Ex ia d IIC T4 Gb Rechargeable
IECEx II 2G Ex ia d IIC T3 Gb Replaceable Batteries
USA
Class I Division 1, Groups A, B, C and D
CSA
Class I Division 1, Groups A, B, C and D
MED (Marine Equipment Directive)
96/98/EC (optional approval)
Safety & Operation Standards
EN50014, EN50020, EN50018, 94/9/EC
EN50270, EN50271, IEC61508
PC Interface and PC Software*
G-TECTA™ PC Software gives the user access to data
and event log files, printing reports, reconfiguring
alarm levels, instrument operation functions, and
running calibrations.
G-TECTA™ 4GP can be connected to a PC by
using the G-TECTA™ 4GP infrared optical link.
The PC requires a G-TECTA™ infrared PC interface and
G-TECTA™ PC software.
Set-up
1. Install G-TECTA™ PC software and G-TECTA™
infrared interface on the PC.
W A R N I N G
The infrared communications are not
IrDA. Do NOT install IrDA drivers.
2. Where applicable, install USB software**
(supplied in USB adaptor box)
3. Attach the infrared RS232 cable to the USB
adaptor (where applicable) and to the PC
4. When using a USB adaptor to connect the
instrument, follow the procedure below:
(i) Before connecting the instrument, check the
available communications ports available on the
G-TECTA™ PC software options screen and note
the ports available
(ii) Switch on the G-TECTA™ 4GP instrument
and align the instrument infrared port with
the infrared adaptor
(iii) Check the available communications ports
available again and select the new port or port 4,
5 or 6 and save your changes by clicking the save
button on the bottom right of the options screen
(iv) Close the G-TECTA™ PC software
Note: If the instrument does not connect to the PC,
select a different communications port and try again
5. Switch on the G-TECTA™ 4GP instrument and
align the instrument infrared port with the
infrared adaptor
6. Open the G-TECTA™ PC software and using either
the Wizard or the Engineer’s Form, select G-TECTA™
4GP and upload the data from the instrument.
7. The software will allow you to access data or
configure the instrument as required.
Note: For more information on using the G-TECTA™
PC software, see the installed help file.