148 Chapter 12
Square wave bolus
Delivers 0.1-25.0 units of insulin over a duration of 30
minutes to eight hours (limited by maximum bolus setting).
Temporary basal rate
Allows the user to temporarily change the current basal rate
for a duration of 30 minutes to 24 hours (limited by maximum
basal setting).
Time and date display
Pump users have the choice of 12-hour or 24-hour formats.
Pump users set the time/date, including the year, month and
day.
The date/time is always displayed on the status screen.
Water tight
Conforms to IEC60601-1 sub clause 44.6 and IEC60529 IPX7
starndard.