28 Chapter 2
Manual Prime
1. Make sure that the infusion set is disconnected from your
pump to your body.
2. Hold ACT to Prime. The pump will prepare to prime and will
beep six times.
If you let go of the ACT button, either hold down ACT again to con-
tinue or press ESC to repeat the entire “Rewind” procedure.
3. Continue to hold the ACT button until you hear a second set of
six beeps.
WARNING: DO NOT continue until you hear the second set of beeps
and the screen from step 4 appears. Call the 24-hour
Product Help-line for assistance.
MiniMed
MANUAL PRIME
Disconnect from Body
Insert/Lock Reservoir
Hold ACT to Prime
MiniMed
PREPARING TO PRIME
Hold ACT