77
Chapter 6
• • • • • •
Suspend
The Suspend function stops all insulin deliveries (Basal, Bolus,
and Fixed Prime) that you have programmed into your pump.
While the pump is in Suspend, you will only be able to view the
STATUS screen. No other menu choices are available.
You must resume your pump in order for the delivery of basal
insulin to restart. Bolus and Fixed Prime will not restart. You
must reprogram and activate the Bolus and Fixed Prime
settings.
MiniMed
MAIN MENU
Bolus
Basal
Prime
Utilities
MiniMed
SUSPEND
Press ACT to Stop Pump
Suspend
ACT