11
Dilution Fresh Air Cal
Please Wait….. 8
The Cannonball 3 will then proceed to
the automatic span calibration.
SPAN CALIBRATION
MODE = ADJUST 5
6. If span calibration is not required,
allow the unit to complete the 5-
second countdown without pressing
any buttons.
Press MODE before the before the end
of the 5-second countdown to initiate the
span calibration sequence. The
Cannonball 3 will then temporarily
disable the dilution pump and show the
following screen:
APPLY GAS
MODE = CANCEL
7a. If using a calibration balloon, fill the
empty balloon with calibration gas
until it is nearly full. Connect the
calibration balloon assembly to the
inlet coupling on the Cannonball 3.
Continue to flow gas to the balloon
throughout the span calibration
procedure.
7b. If using a demand flow regulator,
connect it and flow gas to the
instrument.
Failure to purge the
calibration balloon of all gas prior to
inflation with calibration gas may
result in inaccurate and potentially
dangerous readings.
Make sure the
calibration balloon is filled with the
appropriate calibration gas before
attempting to calibrate the Cannonball
3. Use of improper calibration gas or
failure to nearly fill the calibration
balloon may lead to inaccurate and
potentially dangerous readings.
The Cannonball 3 automatically
recognizes the type of gas supplied and
automatically adjusts the sensors.
The instrument will then show the
following screen as each sensor that may
be calibrated using the current gas
mixture is adjusted in turn. The span
adjustment process from this point on is
automatic and requires no user input.
LEL Sensor Reading 49
CO Sensor Reading 51
H
2
S Sensor Reading 24
Multi Cal Gas Detected
Once the readings stabilize, the
Cannonball will automatically adjust itself
so that the sensor readings match the
expected gas concentration values as
they appear in the gas values
subdirectory of the Calibration Menu.