LEL Calibrated to 50
CO Calibrated to 50
H
2
S Calibrated to 25
The Cannonball 3 will then display the
following screen, which shows the final
calibration values for the current span
calibration, as well as the maximum
adjustment values possible for the LEL
sensor and any toxic sensors that are
currently recognized by the instrument.
As a sensor loses sensitivity, the
maximum possible adjustment will
decrease to approach the expected
concentration of the calibration gas
.
LEL Adjusted to 50
Max Possible to 221
CO Adjusted to 50
Max Possible to 281
H
2
S Adjusted to 25
Max Possible to 138
The instrument will then re-enable the
dilution pump and proceed to the
automatic dilution span calibration.
Calibrating Dilution…..
LEL 50
CO 50
H
2
S 25
Upon completion of the automatic dilution
span calibration, the Cannonball 3 will
proceed through the following screens
before turning itself off.
SPAN CALIBRATION
COMPLETED
BEGIN SHUT DOWN
PLEASE WAIT
Saving Instrument Data
Saving O
2
Sensor Data
Saving LEL Sensor Data
Saving CO Sensor Data
Saving H
2
S Sensor Data
After shut down, immediately remove all
calibration fittings from the Cannonball 3.
Press the MODE button to turn the
instrument on and resume normal
operation.
Note: If the MODE button is pressed
at any time prior to completion of the
calibration, the calibration procedure
will be cancelled and the instrument
will return to normal operation.
Wait at least 3
minutes after turning the instrument
on to allow sensor readings to
stabilize fully before initiating any
calibration procedures. Failure to wait
three minutes before initiating
calibration procedures may lead to
inaccurate and potentially dangerous
readings.
Make sure the
regulator, cylinder seating surfaces
and threads are clean and dry before
attaching the regulator to the cylinder
of gas. Introduction of contaminants
through the regulator fittings may
alter or degrade the concentration of
the gas contained in the cylinder and
may lead to inaccurate and potentially
dangerous gas readings.
Make sure the
calibration balloon is filled with the
appropriate calibration gas before
attempting to calibrate the Cannonball
3. Use of improper calibration gas or
failure to nearly fill the calibration