8
Checking for
Dilution Pump…..
Setting O
2
Alarms
For Dilution Mode
LOW CEIL
O
2
OFF 5.0
MODE=Acknowledge
4. Press the MODE button to
acknowledge the dilution Oxygen
alarms.
Dilution monitoring is
extremely hazardous and should only
be performed by individuals trained in
the safe monitoring of inert
environments. Failure to follow
established procedures while
performing dilution monitoring may
lead to dangerous and potentially life-
threatening conditions.
2.2.2 Shortcut to dilution mode
From the current gas readings screen,
the following shortcut is available for
rapid entry into or out of dilution mode:
O
2
LEL
20.9 0
CO H
2
S
0 0
DILUTION PUMP OFF
Time 6:15 RT 1:04
1. Press the right and left navigation
arrows simultaneously. Release the
buttons when prompted by the
instrument. If enabling the dilution
pump, the following screen will then
be shown.
Turn Dilution Pump On?
YES NO
2. Press the MODE button with YES
highlighted to enable the dilution
pump.
Checking for
Dilution Pump…..
Setting O
2
Alarms
For Dilution Mode
LOW CEIL
O
2
OFF 5.0
MODE=Acknowledge
3. Press the MODE button to
acknowledge the dilution Oxygen
alarms.
Note: This procedure may also be
used to disable the dilution pump.
The Cannonball 3 will then require
acknowledgement of the non-dilution
Oxygen alarm settings.
2.3 Alarm changes in dilution
mode
Changes to the dilution alarm set points
are done in the same way as changes to
the conventional alarm set points.
To change the dilution alarm settings:
1. Enter the Alarms Menu.