9
ALARMS MENU
CUSTOM LATCH
DEFAULT TEMPERATURE
EXIT
2. Use the navigation arrows to highlight
CUSTOM and press the MODE
button. The current alarm settings
screen will then be shown.
CURRENT ALARM SETTINGS
LOW CEIL
O
2
STD
19.5 22.0
O2 DIL
OFF 5.0
LEL 10
CEIL STEL TWA
CO 100 35 100
H
2
S 10 15 10
EXIT
3. Use the up and down navigation
arrows to select O
2
DIL and press the
MODE button.
CURRENT ALARM SETTINGS
LOW CEIL
O
2
DIL
OFF 5.0
4. Use the left and right navigation
arrows to select the low or ceiling
alarm for modification.
Note: The ceiling O
2
dilution alarm is
an ascending alarm. If the amount of
oxygen is higher than the alarm set
point, the visible and audible alarms
will be activated.
CURRENT ALARM SETTINGS
LOW CEIL
O
2
DIL
OFF 6.0
5. Use the up and down navigation
buttons to modify the current setting.
When the alarm adjustment is
completed, press the MODE button
to enter the new setting.
CURRENT ALARM SETTINGS
LOW CEIL
O
2
DIL
OFF 6.0
Save Changes?
YES NO CANCEL
6. Press the MODE button with YES
highlighted to save the new alarm
settings.
3. Calibration in dilution
mode
To ensure O
2
sensor accuracy at levels
approaching 0.0%/Vol. oxygen,
calibration procedures in dilution-capable
instruments include specific sub-routines
involving calibration with 99.9% or better
nitrogen (N
2
).
3.1 Calibration frequency
recommendations: dilution
See Appendix A for Sperian’s calibration
frequency recommendations. Dilution-
capable Cannonball3 instruments should
be calibrated on a similar schedule for
both standard and dilution-related
calibration procedures.
Due to the need for increased accuracy
at levels approaching 0.0%/Vol. oxygen,
it is necessary to perform both the
manual fresh air calibration (see section
3.3.1) and the O
2
zero calibration (see
section 3.3.2) whenever an oxygen